Lo-Ray-Ligh
grade
diffraction
grating
Achieves Ultra-Low Stray Light,
Enabling 8-Abs Measurements
4
Conventional models provide 5-Abs coverage. Even high-end models
have been limited to 6 Abs. In contrast, the UV-2700 achieves ultra-low
stray light levels, expanding the range to 8 Abs, with a transmittance
value of 0.000001 % (1 part in 100 million). This system achieves
high-level absorbance measurements with incomparable precision. In
addition to measuring even high-concentration samples as is, eliminating
the need to dilute samples, the system can be applied to evaluating the
transmission characteristics of polarization films.
Wavelengths in the 400 nm to 650 nm range can be measured to 8 Abs.
Absorbance Linearity UV-2700
Spectral Comparison of Aqueous Potassium Permanganate Solutions
UV-2700
9.0
Calibration Curve
Concentration (mg/L)
This shows the relationship between the absorbance and the
concentration of an aqueous potassium permanganate solution.
Good linearity is evident to 8 Abs.
8.0
6.0
4.0
Abs.
2.0
0.0
0.0 2.0 4.0 6.0 8.0 8.5
This spectrum is measured from an aqueous KMnO4 (potassium permanganate) solution. Unlike conventional models, this system is capable of high-level absorbance measurements.
UV-2700
9.0
8.0
6.0
4.0
Abs.
2.0
0.0
400.0 500.0
nm
600.0 700.0
UV-2550
(Conventional Model)
9.0
8.0
6.0
4.0
Abs.
2.0
0.0
400.0 500.0
nm
600.0 700.0